Guide · EU MDR and IVDR

What are the EU MDR and IVDR transition deadlines for legacy devices?

Under the EU MDR, legacy devices certified under the old directives can be placed on the market until 31 December 2027 (class III, and most class IIb implantables) or 31 December 2028 (other class IIb, class IIa, and class I sterile or measuring devices), provided the manufacturer had an MDR quality system and a notified body application in place by 26 May 2024 and a signed written agreement by 26 September 2024. Under the IVDR, the dates run to the end of 2027, 2028 or 2029 by class, with application deadlines in 2025, 2026 and 2027. The dates only hold while the conditions do.

Last reviewed 19 November 2026 · SMEDTEC

EU MDR: Regulation (EU) 2023/607

Regulation (EU) 2023/607 rewrote the MDR's transitional provisions (Article 120). Devices with a valid certificate under the old directives (93/42/EEC or 90/385/EEC) may be placed on the market or put into service until 31 December 2027 for all class III devices and class IIb implantables (except sutures, staples, dental fillings, dental braces, tooth crowns, screws, wedges, plates, wires, pins, clips and connectors), and until 31 December 2028 for other class IIb devices, class IIa devices, and class I devices that are sterile or have a measuring function.

Devices that needed no notified body under the Medical Devices Directive, had a declaration of conformity drawn up before 26 May 2021, and need a notified body under the MDR may be placed on the market until 31 December 2028.

  • The device continues to comply with the directive it was certified under
  • No significant changes in design or intended purpose
  • No unacceptable risk to health or safety
  • An MDR quality management system in place by 26 May 2024
  • A formal application lodged with a notified body by 26 May 2024, and a written agreement signed with it by 26 September 2024

EU IVDR: Regulation (EU) 2024/1860

Regulation (EU) 2024/1860 amended the IVDR's transitional provisions (Article 110). Devices with a valid certificate under the IVD Directive (98/79/EC) may be placed on the market until 31 December 2027. Devices that needed no notified body under the directive, had a declaration of conformity drawn up before 26 May 2022, and need one under the IVDR may be placed on the market until 31 December 2027 (class D), 31 December 2028 (class C), or 31 December 2029 (class B, and class A devices placed on the market sterile).

The conditions mirror the MDR's: continued compliance with the directive, no significant change, no unacceptable risk, an IVDR quality management system by 26 May 2025, and notified body deadlines that step by class. At the date of this review the class D and class C deadlines have passed; the class B and sterile class A application deadline is 26 May 2027, with the written agreement due by 26 September 2027.

The obligations that already apply

Legacy status covers the conformity assessment, not everything else. Under both regulations, the requirements on post-market surveillance, market surveillance, vigilance and the registration of economic operators and devices apply to legacy devices in place of the directives' (MDR Article 120(3d); IVDR Article 110(3d)). A legacy device needs its PMS plan, its PMS report or PSUR, and its vigilance process now, not at certification.

The European Commission proposed a targeted revision of the MDR and IVDR in December 2025. At the date this page was reviewed it had not been adopted, so the rules below still apply; check whether that has changed before you rely on them.

Tracking each device against its date

The work is per device: which paragraph it falls under, its date, whether each condition is met and evidenced, and where the notified body process stands. In SMEDTEC OS you give each device an EU MDR or EU IVDR conformity track with that date as its target, add work items for the application, the written agreement and the technical documentation, each with an owner and a due date, and keep the evidence for each condition in the document vault. Tracks that fall behind show as at risk on the dashboard. You set the date that applies to each device; SMEDTEC OS does not work it out for you.

Legacy device dates at a glance

Legacy device dates at a glance
RegulationDevicesMay be placed on the market untilNotified body application and agreement
EU MDRClass III; class IIb implantable (with listed exceptions)31 December 202726 May 2024 and 26 September 2024
EU MDROther class IIb; class IIa; class I sterile or measuring31 December 202826 May 2024 and 26 September 2024
EU MDRNo notified body under the directive, needs one under the MDR31 December 202826 May 2024 and 26 September 2024
EU IVDRCertificate under the IVD Directive; or class D31 December 202726 May 2025 and 26 September 2025
EU IVDRClass C31 December 202826 May 2026 and 26 September 2026
EU IVDRClass B; class A sterile31 December 202926 May 2027 and 26 September 2027

A good fit for

  • Manufacturers with legacy devices moving to the MDR or IVDR who need each device's date, conditions and evidence in one place
  • Teams showing a notified body or a distributor where each device stands

Not the right tool for

  • Deciding which paragraph a device falls under or whether a change is significant: that is a regulatory judgement for you and your adviser
  • Writing the technical documentation itself

Questions regulatory teams ask

Common follow-up questions

Read the guides

Is there still a sell-off date for legacy devices already on the market?

No. Regulation (EU) 2023/607 rewrote MDR Article 120(4) and IVDR Article 110(4) so that legacy devices lawfully placed on the market may continue to be made available on the market or put into service, with no end date.

Does SMEDTEC OS tell me which deadline applies to my device?

No. You record the date that applies as the track's target date, and SMEDTEC OS keeps the work, the owners and the evidence against it. The classification and the paragraph a device falls under are your decision.